Instrument Inputs/Outputs | |
USB ports | USB On-The-Go (OTG) |
RS-232 port | Yes |
Video output | VGA output standard |
Analog output | 1 analog output (optional), Selectable 1 V/10 V Full Scale, 4 mA max |
Digital output | 3 alarm outputs, 5V TTL, 10 mA |
Environmental Ratings | |
IP rating | IP rating Designed to meet the standards of the Ingress Protection (IP) rating number IP67 (navigation pad version) or IP66 (knob version) per IEC 60529-2004 (Degrees of Protection provided by enclosures - IP Code). The product design was confirmed to meet the IP rating by means of Olympus NDT's internal design verification test process that occurs prior to the release of the product to production. |
Explosive atmosphere | Safe operation as defined by Class I, Division 2, Group D, as defined in the National Fire Protection Association Code (NFPA 70), Article 500, and tested using MIL-STD-810F, Method 511.4, Procedure I. |
Shock tested | MIL-STD-810F, Method 516.5 Procedure I, 6 cycles each axis, 15 g, 11 ms half sine. |
Vibration tested | MIL-STD-810F, Method 514.5, Procedure I, Annex C, Figure 6, general exposure: 1 hour each axis |
Operating temperature | -10 °C to 50 °C (14 °F to 122 °F) |
Battery storage temperature | 0 °C to 50 °C (32 °F to 122 °F) |
General | |
Overall dimensions (W x H x D) | 236 mm x 167 mm x 70 mm (9.3 in. x 6.57 in. x 2.76 in.) |
Weight | 1.68 kg (3.72 lb), including lithium-ion battery |
Keypad | English, International, Japanese, Chinese |
Languages | English, Spanish, French, German, Japanese, Chinese, Portuguese |
Transducer connections | BNC or Number 1 LEMO |
Data storage | 50,000 IDs onboard, removable 2 GB MicroSD card (standard) |
Battery type | Single lithium-ion rechargeable standard |
Battery life | 12 h to 13 h (lithium-ion), 3 h (alkaline) |
Power requirements | AC Mains: 100 VAC to 120 VAC, 200 VAC to 240 VAC, 50 HZ to 60 Hz |
Display type | Full VGA (640 x 480 pixels) transflective color LCD, 60 Hz update rate |
Display dimensions (W x H, Diag.) | 117 mm x 89 mm, 146 mm (4.62 in. x 3.49 in., 5.76 in.) |
Pulser | |
Pulser | Tunable Square Wave |
PRF | 10 Hz to 2000 Hz in 10 Hz increments |
Energy settings | 100 V, 200 V, 300 V or 400 V |
Pulse width | Adjustable from 25 ns to 5,000 ns (0.1 MHz) with PerfectSquare™ technology |
Damping | 50, 100, 200, 400 Ω |
Receiver | |
Gain | 0 to 110 dB |
Maximum input signal | 20 V p-p |
Receiver input impedance | 400 Ω ± 5% |
Receiver bandwidth | 0.2 MHz to 26.5 MHz at -3 dB |
Digital filter settings | Eight digital filter sets standard (0.2-10 MHz, 2.0-21.5 MHz, 8.0-26.5 MHz, 0.5-4 MHz, 0.2-1.2 MHz, 1.5-8.5 MHz, 5-15 MHz, DC-10 MHz) |
Rectification | Full-wave, Positive Half-wave, Negative Half-wave, RF |
System linearity | Horizontal: ± 0.5% FSW |
Resolution | 0.25% FSH, amplifier accuracy ± 1dB |
Reject | 0 to 80% FSH with Visual Warning |
Amplitude measurement | 0 to 110% full screen height with 0.25% resolution |
Measurement rate | Equivalent to PRF in all modes |
Calibration | |
Automated calibration | Velocity, Zero Offset Straight Beam (First Backwall or Echo-to-Echo) Angle Beam (Soundpath or Depth) |
Test modes | Pulse Echo, Dual, or Through Transmission |
Units | Millimeters, inches, or microseconds |
Range | 3.36 mm to 13,405 mm (0.132 in. to 527 in.) @ 5,900 m/s (0.2320 in./μ) |
Velocity | 635 m/s to 15240 m/s (0.0250 in./µs to 0.6000 in./µs) |
Zero offset | 0 µs to 750 µs |
Display delay | -59 mm to 13401 mm (-2.323 in. to 527.6 in.) @ longitudinal velocity in steel |
Refracted angle | 0° to 85° in 0.1° increments |
Gates | |
Measurement gates | 2 fully independent gates for amplitude and TOF measurements |
Gate start | Variable over entire displayed range |
Gate width | Variable from Gate Start to end of displayed range |
Gate height | Variable from 2 to 95% full screen height |
Alarms | Positive and Negative Threshold, Minimum Depth (Gate 1 and Gate 2) |
Measurements | |
Measurement display locations | 5 locations available (manual or auto selection) |
Gate 1 | Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight, Min./Max. Depth, Min./Max. Amplitude |
Gate 2 | Same as Gate 1 |
Echo-to-Echo | Standard Gate 2-Gate 1, Optional IF Gate Tracking |
Other measurements | Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 rating (D), Reject Value |
DAC/TVG | Standard |
DAC points | Up to 50 points, 110 dB dynamic range |
Special DAC modes | Custom DAC (up to 6 curves) |
Curved surface correction | Standard OD or Bar correction for Angle Beam measurements |